Database last updated on Sun, 13 Sep 2026 18:13:06 +0000 by localhost

Critical warnings from b-files' comments for days between 2019086 and 2019086 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 8791B08619.026TORON#02611:45:451-1
hg:Lamp not on... hg test terminated F1= 0 F4= 8791B08619.026TORON#02611:45:451-1

Instrument constant changes for the week before 2019086 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 25 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
026201908611919.0

Non-critical warnings from b-files' comments for days between 2019086 and 2019086 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #1 reference found at step-797 (+CAL)= 122 Now set to o3 offset = 2936199B08619.026TORON#02600:57:1112210009000
sc:SC intensity too low ( 1/cy), skipping.23B08619.026TORON#02602:37:261-1
ds:Suppressed oscillation of filter wheel.15B08619.026TORON#02619:53:281-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
086TORONTO#026ecsc69692110433265723
(0 sec)

Relative and absolute humidity for day 2019086 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
02610.01.7