Database last updated on Sun, 13 Sep 2026 01:41:14 +0000 by localhost

Critical warnings from b-files' comments for days between 2019076 and 2019076 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 2191B07619.026TORON#02611:52:151-1
hg:Lamp not on... hg test terminated F1= 0 F4= 2191B07619.026TORON#02611:52:151-1

Instrument constant changes for the week before 2019076 (1 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 24 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0262019076817.5

Non-critical warnings from b-files' comments for days between 2019076 and 2019076 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #1 reference found at step-40 (+CAL)= 879 Now set to o3 offset = 2936299B07619.026TORON#02611:53:2887910009000
ds:DS intensity too high ( 225744/cy) quitting.399B07619.026TORON#02615:30:391-1
ds:Suppressed oscillation of filter wheel.35B07619.026TORON#02617:47:541-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
076TORONTO#026ecsc61612137221265633
(0 sec)

Relative and absolute humidity for day 2019076 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
02610.41.7