Critical warnings from b-files' comments for days between 2019068 and 2019068 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 2351B06819.026TORON#02613:48:081-1
hg:Lamp not on... hg test terminated F1= 0 F4= 2351B06819.026TORON#02613:48:081-1
ds:Suppressed oscillation of filter wheel.55B06819.026TORON#02615:40:591-1

Instrument constant changes for the week before 2019068 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 23 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0262019068923.1

Non-critical warnings from b-files' comments for days between 2019068 and 2019068 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #1 reference found at step-867 (+CAL)= 52 Now set to o3 offset = 2936199B06819.026TORON#02613:49:225210009000

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
068TORONTO#026ecsc808021428265419
(0 sec)

New operators? (0 sec, 91 rows)

OperatorSerial
Time026
(CUT)026
(HH:MM)026

Relative and absolute humidity for day 2019068 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0269.81.6