Database last updated on Sat, 12 Sep 2026 22:40:57 +0000 by localhost

Critical warnings from b-files' comments for days between 2019059 and 2019059 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 1 f4= 2271B05919.026TORON#02616:49:131-1
hg:Lamp not on... hg test terminated F1= 1 F4= 2271B05919.026TORON#02616:49:131-1

Instrument constant changes for the week before 2019059 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 24 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0262019059821.4

Non-critical warnings from b-files' comments for days between 2019059 and 2019059 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:HOME key pressed199B05919.026TORON#02614:31:461-1
ds:Going to menu199B05919.026TORON#02614:31:481-1
si:New SI, HC = -4, NC = 250 Old = -4 , 235199B05919.026TORON#02614:35:241-1
fr:Micrometer #1 reference found at step-917 (+CAL)= 2 Now set to o3 offset = 2909299B05919.026TORON#02616:50:23210009000
ds:Suppressed oscillation of filter wheel.15B05919.026TORON#02618:07:391-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsedhgsluvzs
059TORONTO#026menu585812054517
(0 sec)

Relative and absolute humidity for day 2019059 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
02617.21.7