Database last updated on Sun, 13 Sep 2026 13:40:55 +0000 by localhost

Critical warnings from b-files' comments for days between 2019052 and 2019052 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 4811B05219.026TORON#02617:52:481-1
hg:Lamp not on... hg test terminated F1= 0 F4= 4811B05219.026TORON#02617:52:481-1

Instrument constant changes for the week before 2019052 (4 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 14 rows)

HG burned out? (AP results when HG is ON) (0 sec, 5 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0262019052425.7

Non-critical warnings from b-files' comments for days between 2019052 and 2019052 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:Suppressed oscillation of filter wheel.15B05219.026TORON#02619:07:101-1
ds:DS intensity too high ( 266948/cy) quitting.399B05219.026TORON#02619:11:431-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
052TORONTO#026ecsc10112135114255463
(0 sec)

Relative and absolute humidity for day 2019052 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
02610.01.3