Database last updated on Sun, 13 Sep 2026 07:11:07 +0000 by localhost

Critical warnings from b-files' comments for days between 2019022 and 2019022 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not detected f1= 0 f4= 26841B02219.026TORON#02615:52:111-1
hg:Lamp not on... hg test terminated F1= 0 F4= 26841B02219.026TORON#02615:52:111-1

Instrument constant changes for the week before 2019022 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 8 rows)

HG burned out? (AP results when HG is ON) (0 sec, 11 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0262019022828.0

Non-critical warnings from b-files' comments for days between 2019022 and 2019022 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
zs:ZS intensity too low ( 279/cy), skipping.199B02219.026TORON#02613:05:101-1
fr:Micrometer #1 reference found at step-1154 (+CAL)=-235 Now set to o3 offset = 2909499B02219.026TORON#02615:53:29(+CAL)=-23510009000

Measurements and tests summary.


_jday_loc_serial_skcdsdtedfmfzhgrssluvzs
022TORONTO#026ecsc3221433192153650
(0 sec)

Relative and absolute humidity for day 2019022 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0267.00.8