Database last updated on Mon, 14 Sep 2026 14:11:05 +0000 by localhost

Critical warnings from b-files' comments for days between 2018164 and 2018164 (1 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 1 F4= 861751B16418.026CHURC#02601:49:521-1
hg:Lamp not detected f1= 1 f4= 861721B16418.026CHURC#02601:49:531-1
hg:HG failed841B16418.026CHURC#02603:53:06failed1-1

Instrument constant changes for the week before 2018164 (2 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0262018164321.1

Non-critical warnings from b-files' comments for days between 2018164 and 2018164 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #1 reference found at step-1004 (+CAL)=-85 Now set to o3 offset = 2937199B16418.026CHURC#02603:22:51(+CAL)=-8510009000
hg:HOME key pressed199B16418.026CHURC#02612:12:241-1
hg:Going to menu199B16418.026CHURC#02612:12:261-1
ze:Zenith discrepancy = -2571199B16418.026CHURC#02612:14:01-2571-1010

Measurements and tests summary.


_jday_loc_serial_skcaodedshgsluvzs
164CHURCHILL#026ecsc441011721
(0 sec)

New operators? (1 sec, 170 rows)

OperatorSerial
Time026
(CUT)026
(HH:MM)026
10:40026
11:01026
11:02026
11:23026

Relative and absolute humidity for day 2018164 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0267.40.9