Database last updated on Tue, 15 Sep 2026 16:41:53 +0000 by localhost

Critical warnings from b-files' comments for days between 2018080 and 2018080 (1 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:No intensity detected. Check PMT.81B08018.026CHURC#02601:24:221-1
sl:Lamp not on... sl test terminated F1= 0 F4= 4253101B08018.026CHURC#02614:18:251-1
hg:Lamp not detected f1= 0 f4= 13081B08018.026CHURC#02615:18:561-1
hg:HG failed1541B08018.026CHURC#02615:22:38failed1-1

Instrument constant changes for the week before 2018080 (4 sec, 0 rows)

HG burned out? (AP results when HG is ON) (0 sec, 24 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
026201808027.9

Non-critical warnings from b-files' comments for days between 2018080 and 2018080 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
re:Zenith discrepancy = -2291199B08018.026CHURC#02606:25:19-2291-1010
ds:DS intensity too low ( 1/cy), skipping.399B08018.026CHURC#02612:06:541-1
zs:ZS intensity too low ( 15/cy), skipping.499B08018.026CHURC#02612:07:091-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
080CHURCHILL#026ecsc1414215921147
(0 sec)

New operators? (0 sec, 43 rows)

Relative and absolute humidity for day 2018080 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
02615.23.8