| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
|---|---|---|---|---|---|---|---|---|---|---|
| re: | Previous RE didn't seem to help, so not doing reset again. Operator intervention required. Quitting to menu. | 1 | 1 | B23819.019 | ALERT | #019 | 01:49:41 | 1 | -1 | |
| hg: | Lamp not on... hg test terminated F1= 0 F4= 435 | 11 | 1 | B23819.019 | ALERT | #019 | 11:52:11 | 1 | -1 | |
| hg: | Lamp not detected f1= 0 f4= 435 | 11 | 1 | B23819.019 | ALERT | #019 | 11:52:12 | 1 | -1 | |
| hg: | HG failed | 5 | 1 | B23819.019 | ALERT | #019 | 12:52:55 | failed | 1 | -1 |
| Serial | YYYYJJJ | HG large | max correction |
|---|---|---|---|
| 019 | 2019238 | 4 | 5.7 |
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
|---|---|---|---|---|---|---|---|---|---|---|
| re: | Going to menu | 1 | 99 | B23819.019 | ALERT | #019 | 01:49:42 | 1 | -1 |
| _jday | _loc | _serial | _skc | aode | ds | hg | sl | uv | zs |
|---|---|---|---|---|---|---|---|---|---|
| 238 | ALERT | #019 | menu | 111 | 111 | 8 | 6 | 55 | 17 |
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
|---|---|---|---|---|
| 019 | 9.8 | 1.1 |