Database last updated on Fri, 11 Sep 2026 00:40:59 +0000 by localhost

Critical warnings from b-files' comments for days between 2020056 and 2020056 (0 sec, 4 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Lamp not on... hg test terminated F1= 1 F4= 021B05620.018GOOSE#01804:43:281-1
hg:Lamp not detected f1= 1 f4= 021B05620.018GOOSE#01804:43:291-1
hg:HG failed11B05620.018GOOSE#01804:43:30failed1-1
hg:No intensity detected. Check PMT.11B05620.018GOOSE#01811:49:311-1

Instrument constant changes for the week before 2020056 (2 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 22 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
018202005623.2

Non-critical warnings from b-files' comments for days between 2020056 and 2020056 (1 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
ds:DS intensity too high ( 241259/cy) quitting.199B05620.018GOOSE#01817:32:021-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedhgrssluvzs
056GOOSE BAY#018ecsc47472113264739
(1 sec)

Relative and absolute humidity for day 2020056 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0184.60.4