Critical warnings from b-files' comments for days between 2025292 and 2025292 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
sc:SC intensity too low ( 107/cy), skipping.73B29225.017TORON#01713:39:431-1

Instrument constant changes for the week before 2025292 (2 sec, 0 rows)

Non-critical warnings from b-files' comments for days between 2025292 and 2025292 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
dp:Doing dispersion with internal HG.199B29225.017TORON#01708:08:201-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfzhgrssluvzs
292TORONTO#017calsc,Next schedule will be calsc (now calsc ZS= 90 ZS= 90.542),Next schedule will be calsc (now calsc ZS= 90 ZS= 90.738),Waiting until ZA = -90 now -100.7,Waiting until ZA = -90 now -102.5,Waiting until ZA = -90 now -104.3,Waiting until ZA = -90 now -106.1,Waiting until ZA = -90 now -107.9,Waiting until ZA = -90 now -109.7,Waiting until ZA = -90 now -111.5,Waiting until ZA = -90 now -113.3,Waiting until ZA = -90 now -115.1,Waiting until ZA = -90 now -116.9,Waiting until ZA = -90 now -118.7,Waiting until ZA = -90 now -120.4,Waiting until ZA = -90 now -122.1,Waiting until ZA = -90 now -123.9,Waiting until ZA = -90 now -91.7,Waiting until ZA = -90 now -93.5,Waiting until ZA = -90 now -95.3,Waiting until ZA = -90 now -97.1,Waiting until ZA = -90 now -98.9919122523261247
(1 sec)

Number of days since last dome/window cleaning (0 sec, 2 rows)

serialdays
017more than a year

Missing files (0 sec, 1 rows)

Seriallast datadays ago
01720252930-1

Questionable Run/Stop results. (0 sec, 1 rows)

SerialJdayslit 0dark rsdark countsslit 1slit 2slit 3slit 4slit 5slits 3+5
01720252920.9981.0003 / 3 (1.8M)0.9950.9990.9960.9990.9990.993

Run/Stop results for last week. (0 sec, 1 rows)

SerialTotalFailed 0Failed darkFailed 1Failed 2Failed 3Failed 4Failed 5Failed double
0171406101003

AP results for MB Brewer. (0 sec, 1 rows)

SerialJdayt1 offt1 ont2 offt2 ont3 offt3 onHV offHV on+15V off+15V on+5V off+5V on-15V off-15V on+24V off+24V on+5V sps off+5V sps on-8V sps off-8V sps onSL A offSL A onSL V offSL V onW
017202529222.222.324.524.522.022.01424.61426.415.014.35.05.0-14.9-14.924.124.34.94.9-8.0-8.04.61.60.110.6716.6

Standard lamp intensity analysis (0 sec, 1 rows)

Serialminmaxfilterfixable30-day history
#017213651021369800yes