Database last updated on Sun, 13 Sep 2026 18:13:06 +0000 by localhost
Critical warnings from b-files' comments for days between 2019121 and 2019121 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| hg: | Lamp not on... hg test terminated F1= 3 F4= 9 | 1 | 1 | B12119.013 | RESOL | #013 | 08:48:51 | | 1 | -1 |
| hg: | Lamp not detected f1= 3 f4= 9 | 1 | 1 | B12119.013 | RESOL | #013 | 08:48:52 | | 1 | -1 |
Instrument constant changes for the week before 2019121 (6 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 32 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 013 | 2019121 | 4 | 11.6 |
Non-critical warnings from b-files' comments for days between 2019121 and 2019121 (0 sec, 2 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| fr: | Micrometer #1 reference found at step-379 (+CAL)= 547 Now set to o3 offset = 1700 | 1 | 99 | B12119.013 | RESOL | #013 | 08:49:59 | 547 | 1000 | 9000 |
| ds: | Suppressed oscillation of filter wheel. | 1 | 5 | B12119.013 | RESOL | #013 | 17:43:29 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fz | hg | rs | sl | uv | zs |
| 121 | RESOLUTE | #013 | ecsc | 109 | 109 | 1 | 1 | 17 | 30 | 1 | 6 | 92 | 48 |
(0 sec)
Relative and absolute humidity for day 2019121 (3 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 013 | 10.4 | 1.9 |