Database last updated on Sun, 13 Sep 2026 22:11:19 +0000 by localhost
Critical warnings from b-files' comments for days between 2018257 and 2018257 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| hg: | Lamp not detected f1= 2 f4= 298 | 3 | 1 | B25718.011 | TORON | #011 | 05:31:04 | | 1 | -1 |
| hg: | Lamp not on... hg test terminated F1= 2 F4= 298 | 3 | 1 | B25718.011 | TORON | #011 | 05:31:04 | | 1 | -1 |
| hg: | HG failed | 2 | 1 | B25718.011 | TORON | #011 | 05:31:05 | failed | 1 | -1 |
Instrument constant changes for the week before 2018257 (3 sec, 0 rows)
SL burned out? (AP results when SL is ON) (0 sec, 2 rows)
Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)
| Serial | YYYYJJJ | HG large | max correction |
| 011 | 2018257 | 4 | 98.0 |
Non-critical warnings from b-files' comments for days between 2018257 and 2018257 (0 sec, 3 rows)
| Type | Comment | count | priority | fname | location | serial | time | value | min | max |
| sc: | SC intensity too low ( 3/cy), skipping. | 1 | 3 | B25718.011 | TORON | #011 | 04:56:17 | | 1 | -1 |
| sc: | Suppressed oscillation of filter wheel. | 2 | 5 | B25718.011 | TORON | #011 | 13:23:15 | | 1 | -1 |
| ds: | DS intensity too high ( 200005/cy) quitting. | 1 | 99 | B25718.011 | TORON | #011 | 20:40:55 | | 1 | -1 |
Measurements and tests summary.
| _jday | _loc | _serial | _skc | aode | ds | dt | ed | fz | hg | rs | sl | zs |
| 257 | TORONTO | #011 | calsc | 156 | 157 | 2 | 1 | 4 | 40 | 2 | 7 | 3 |
(1 sec)
New operators? (0 sec, 173 rows)
Relative and absolute humidity for day 2018257 (1 sec, 1 rows)
| Serial | RH, % | AH, g/m3 | 7-day RH history | Des-nt replcd days ago |
| 011 | 8.6 | 4.8 |