Database last updated on Mon, 14 Sep 2026 05:11:04 +0000 by localhost

Critical warnings from b-files' comments for days between 2018172 and 2018172 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:No intensity detected. Check PMT.21B17218.011TORON#01106:00:541-1
ds:Suppressed oscillation of filter wheel.75B17218.011TORON#01112:48:171-1

Instrument constant changes for the week before 2018172 (3 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0112018172327.7

Non-critical warnings from b-files' comments for days between 2018172 and 2018172 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #1 reference found at step 11029 (+CAL)= 12365 Now set to o3 offset = 12404199B17218.011TORON#01105:32:191236510009000
ds:DS intensity too high ( 300929/cy) quitting.299B17218.011TORON#01113:38:151-1
sc:SC intensity too low ( 48/cy), skipping.13B17218.011TORON#01123:06:431-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
172TORONTO#011calsc1421432119639273514
(0 sec)

New operators? (0 sec, 46 rows)

Relative and absolute humidity for day 2018172 (2 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0115.82.9