Database last updated on Mon, 14 Sep 2026 07:11:02 +0000 by localhost

Critical warnings from b-files' comments for days between 2018143 and 2018143 (0 sec, 1 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:No intensity detected. Check PMT.31B14318.011TORON#01107:23:171-1

Instrument constant changes for the week before 2018143 (3 sec, 0 rows)

SL burned out? (AP results when SL is ON) (0 sec, 8 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
0112018143332.0

Non-critical warnings from b-files' comments for days between 2018143 and 2018143 (0 sec, 2 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
fr:Micrometer #1 reference found at step 11021 (+CAL)= 12357 Now set to o3 offset = 12404199B14318.011TORON#01105:27:531235710009000
ds:Suppressed oscillation of filter wheel.15B14318.011TORON#01113:25:241-1

Measurements and tests summary.


_jday_loc_serial_skcaodedsdtedfmfzhgrssluvzs
143TORONTO#011calsc14114121211041273315
(0 sec)

New operators? (0 sec, 9 rows)

Relative and absolute humidity for day 2018143 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
01113.25.8