Database last updated on Tue, 15 Sep 2026 02:11:06 +0000 by localhost

Critical warnings from b-files' comments for days between 2017352 and 2017352 (0 sec, 5 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:HG cal too far off limits - resetting micrometer(s)31B35217.011TORON#01113:00:501-1
hg:Lamp not detected f1= 0 f4= 37051B35217.011TORON#01113:19:451-1
hg:Lamp not on... hg test terminated F1= 0 F4= 37051B35217.011TORON#01113:19:451-1
hg:*** Micrometer reset attempted - HG cal still failed12B35217.011TORON#01113:46:421-1
hg:HG failed11B35217.011TORON#01113:46:42failed1-1

Instrument constant changes for the week before 2017352 (1 sec, 0 rows)

Hg with step corrections larger than 10 steps or multiple corrections larger than 2.5 steps (0 sec, 1 rows)

SerialYYYYJJJHG largemax correction
01120173522521188.0

Non-critical warnings from b-files' comments for days between 2017352 and 2017352 (0 sec, 3 rows)

TypeCommentcountpriorityfnamelocationserialtimevalueminmax
hg:Going to menu199B35217.011TORON#01113:46:431-1
hg:DEL key pressed499B35217.011TORON#01114:35:471-1
hs:HOME key pressed299B35217.011TORON#01116:05:531-1

Measurements and tests summary.


_jday_loc_serial_skchg
352TORONTO#011menu34
(1 sec)

New operators? (0 sec, 73 rows)

Relative and absolute humidity for day 2017352 (1 sec, 1 rows)

SerialRH, %AH, g/m37-day RH historyDes-nt replcd days ago
0117.02.4